{"id":"CVE-2026-66875","aliases":[],"url":"https://o3.security/vulnerability/CVE-2026-66875","summary":"In the Mira hormone monitor device firmware v1.7.1.47 build 01070147, a remote unauthenticated attacker within BLE range (approximately 10–30 meters) can silently rebind the device…","details":"In the Mira hormone monitor device firmware v1.7.1.47 build 01070147, a remote unauthenticated attacker within BLE range (approximately 10–30 meters) can silently rebind the device to an attacker-controlled account, extract stored hormone measurements in cleartext, cause a denial-of-service via malformed or undocumented command opcodes, and passively track the user via a static random BLE address that never rotates.","published":"2026-08-11T22:18:54.587","modified":"2026-08-11T22:18:54.587","cvss":{"score":8.8,"severity":"HIGH","vector":"CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H"},"epss":null,"cisaKev":null,"exploitsKnown":null,"affectedPackages":[],"fix":null,"references":[{"type":"WEB","url":"https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2026/icsma-26-223-01.json"},{"type":"WEB","url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-26-223-01"}],"provenance":{"sources":["OSV.dev","NVD","FIRST.org (EPSS)"],"lastVerified":"2026-08-11T22:18:54.587"}}